Instrument Calibration Protocol for Alpine Air Jet Sieve
INSTRUMENT | Alpine Air Jet Sieve 200 LS-N. |
Instrument Code No. | |
Refer SOP No. | |
Model No. | 200 LS-N |
Make | Hosokawa |
Date of Calibration | |
Next Due Date |
1. Calibration of Balance
CALIBRATION AGAINST STANDARD WEIGHTS
Standard weight ID:
Weight | 1 g ( ) | 5 g ( ) | 10 g.( ) | 50 g ( ) | 100 g ( ) | 200 g ( ) |
Tolerance | ± 0.1% ( – ) | ± 0.1% ( – ) | ± 0.1% ( – ) | ± 0.1% ( – ) | ± 0.1% ( – ) | ± 0.1% ( – ) |
Observation |
2. Test For Precision:
Take the weight of the same sample six times and calculate the % RSD for particle size analysis.
Details of Material used for the Calibration:
Material Name | |
Batch No. | |
Insp. Lot No. / A.R. No. | |
Mfg. Date | |
Exp. Date |
Sr. No. | Weight of Sample (g) | Sample passes through 250 mm |
1 | ||
2 | ||
3 | ||
4 | ||
5 | ||
6 |
Average
SD
%RSD
Limit: NMT 10.0 %
3. Calibration of Time :
[Stopwatch___________________]
Sr. No. | Set Time (Second) | Displayed Time (Second) | Actual Time in stopwatch (Second) | Difference (Second) | Remarks |
1 | |||||
2 | |||||
3 |
Limit: The difference between time displayed on the instrument and that on the stopwatch should not be more than ± 5 seconds.
Remarks: Calibration Satisfactory / Not Satisfactory
— | Calibrated By | Checked By | Approved By |
Signature | |||
Date | |||
Name | |||
Department | Quality Control | Quality Control | Quality Control |