Instrument Calibration Protocol for Analytical Balance
| Instrument | Analytical Balance |
| Instrument ID. Number | |
| Refer SOP No. | |
| Model No. | |
| Make | |
| Date of Calibration | |
| Next due date |
Weight Box ID No. :
1. Full range and Accuracy Test :
Observation:
| Test Weight (mg/g) | Mass Value of Test Wt. (mg/g) | Maximum Weight (mg/g) | Minimum Weight (mg/g) | Observed Weight (mg/g) |
| 20 mg | ||||
| 50 mg | ||||
| 200 mg | ||||
| 500 mg | ||||
| 2 gm | ||||
| 10 gm | ||||
| 20 gm | ||||
| 50 gm | ||||
| 100 gm | ||||
| 200 gm |
Permissible Tolerance Limit: NMT ± 0.10 %
Remarks: Complies / Does Not Complies.
Calibrated by: Checked by:
2. Repeatability Check :
Observation:
Test Weight: 50 mg
| Sr. No. | Displayed Weight |
| 1 | |
| 2 | |
| 3 | |
| 4 | |
| 5 | |
| 6 | |
| 7 | |
| 8 | |
| 9 | |
| 10 |
Permissible Tolerance: NMT ± 0.10%
Standard Deviation (S.D): ___________
Calculated Repeatability: ___________
Calculation:
Compare S.D with 0.41d, where d = scale interval (d = 0.01 mg, For 0 to 81 g) and (d = 0.1 mg, For 81 to 220 g).
Standard Deviation:
a. If Standard deviation < 0.41d
Repeatability = (2*0.41d) / desired smallest net weight
b. If Standard deviation < 0.41d
Repeatability = (2*S.D) / desired smallest net weight
Calculation:
Minimum Weight Calculation:
Minimum Weight = 2000*S.D.
Calculation:
Remarks: Complies / Does Not Complies.
Calibrated by: Checked by:
3. Eccentricity Test:
Observation:
Test Weight: 100 gm
Permissible Deviation from the center: NMT ±0.05%
| Reference Position | Displayed Weight | Deviation from the Center (C – Displayed Weight) |
| Center (C) | ||
| C1 | ||
| C2 | ||
| C3 | ||
| C4 |
Calculation:
For C,
= Deviation from the Center * 100/Displayed Weight
=
For C1,
= Deviation from the Center * 100/Displayed Weight
=
For C2,
= Deviation from the Center * 100/Displayed Weight
=
For C3,
= Deviation from the Center * 100/Displayed Weight
=
For C4,
= Deviation from the Center * 100/Displayed Weight
=
Acceptance Criteria: Tolerance Limit = NMT ± 0.05 %.
Remarks: Complies / Does Not Complies.
Calibrated by: Checked by:
4. Linearity Test:
Observation:
| Test Weight (mg/g) | Mass value of Test Weight (mg/g) | Observed weight (mg/g) | Deviation (mg/g) | Tolerance Limit : NMT ± 0.05 % |
| 20 gm | ||||
| 50 gm | ||||
| 100 gm | ||||
| 200 gm |
Remarks: Complies/ Does not Complies.
Calibrated by: Checked by:
5. Sensitivity Test:
Observation:
Sensitivity = (Change in Weighing / Change in Load) * 100
=
Deviation = Sensitivity – 100
Deviation =
Acceptance Criteria: Sensitivity NMT ± 0.05 %.
Remarks: Complies/ Does not Complies.
Calibrated by: Checked by:
Remarks: Calibration Satisfactory / Not Satisfactory
| — | Calibrated By | Checked By | Approved By |
| Signature | |||
| Date | |||
| Name | |||
| Department | Quality Control | Quality Control | Quality Control |