Calibration protocol for Tapped Density Tester                  

Instrument calibration protocol for Tapped Density Tester                  

InstrumentTapped Density Tester
Instrument Code Number 
Reference SOP No. 
MakeElectrolab
Model No.ETD-1020
Date of Calibration 
Next Due Date 

For USP I:

Calibration of Strokes / Minute:

Calibration Stop Watch Code: __________________ Valid up to: ____________________

Number of Stokes / Minute 1st ObservationNumber of Stokes / Minute 2nd ObservationNumber of Stokes / Minute 3rd ObservationAverageLimits: 300 + 3 Strokes / Minute
     

For USP II:

Calibration of Strokes / Minute:

Calibration Stop Watch Code: __________________ Valid up to: ____________________

Number of Stokes / Minute 1st ObservationNumber of Stokes / Minute 2nd ObservationNumber of Stokes / Minute 3rd ObservationAverageLimits: 250 + 3 Strokes / Minute
     

Remarks: Calibration Satisfactory / Not Satisfactory

Calibrated By (Sign/ Date)

Checked By (Sign/ Date)