Calibration Protocol for temperature and humidity chambers

Instrument Calibration Protocol for temperature and humidity chambers

INSTRUMENT 
Refer SOP No. 
Model No.                         
Instrument ID. No.               
FrequencyHalf yearly “-15 days”
Date of Calibration 
Next Due Date 

Condition: 25°C/60% RH, 40°C/75% RH, 30°C/65% RH, 30°C/75% RH, 25°C/40%

RH &40°C/20% RH

1) Accuracy:

  • Place appropriate number of calibrator sensor of Temperature & RH near Temperature & RH sensor (Enviro/Master sensor) of stability chamber based on the number of sensors present in

       the stability chamber.

  • Record the Temperature & RH for accuracy study at 5 minutes intervals for the time duration specified for 1 hour.
  • Set Temperature & RH: _______________

Calibrator:

   RecordTemperature (ºC)Relative Humidity (% RH)
Average    

Stability Chamber:

   RecordTemperature (ºC)Relative Humidity (% RH)
Average    
DifferenceTemperature (ºC) (± 0.2 ºC)Relative Humidity (% RH) (±2 % RH)
     

2) Verification:

  • Place 1 calibrator sensor of Temperature & RH nearthe eachTemperature & RH sensor of stability chamber.
  • Record the Temperature & RH for 5 minutes for the time duration specified for 1 hour.

Calibrator:

Record Average Temperature and Relative Humidity

ChannelsTemperature (ºC)Relative Humidity (% RH)
CH-1  
CH-2  
CH-3  
CH-4  
CH-5  
CH-6  
CH-7  
CH-8  
CH-9  
CH-10  

Stability Chamber

Record Average Temperature and Relative Humidity

ChannelsTemperature (ºC)Relative Humidity (% RH)
CH-1  
CH-2  
CH-3  
CH-4  
CH-5  
CH-6  
CH-7  
CH-8  
CH-9  
CH-10  

Limits for Difference:

Temperature (ºC): ± 0.5 ºC

Relative Humidity (% RH): ±2 % RH

ChannelsTemperature Difference (ºC)Relative Humidity Difference (% RH)
CH-1  
CH-2  
CH-3  
CH-4  
CH-5  
CH-6  
CH-7  
CH-8  
CH-9  
CH-10  

Acceptance Criteria:

1) For Accuracy: Difference of the temperature & RH between calibrator and actual temperature of chamber should be within the Specified Limit of ± 0.2° C & ± 2 % respectively

2) For Verification: Difference between Temperature & RH of calibrator & Temperature & RH sensor should be within the specified limits of ± 0.5°C & ± 2 % RH respectively.

Conclusion: Calibration found satisfactory/ not satisfactory.

Calibrated By (Sign/ Date)

Checked By (Sign/ Date)