Instrument Calibration Protocol of Auto Titrator (905 titrando)
INSTRUMENT | Auto Titrator (905 titrando) |
Refer SOP No. | |
Model No. | 905 titrando |
Instrument Code No. | |
Make | Metrohm |
Frequency | Quarterly “- 5 days” |
Date of Calibration | |
Next Due Date |
1. Calibration of pH glass electrode (Aqueous) :
[ Electrode ID :______________________ ]
Date | Buffer | Potential value | Slope | pH of buffer 7.00 | Acceptance criteria |
Slope should between 96-102% Buffer 7.00 pH – should between 6.75-7.25 | |||||
— | — | — | — | ||
— | — | — | — |
Conclusion: Calibration of pH glass electrode (Aqueous)Complies/Does not complies as per
Acceptance criteria.
2. Calibration of Non – aqueous electrode (LL Solvotrode) :
[Electrode ID: ______________________]
Date | Buffer | Potential value | Potential Difference | Acceptance criteria |
Potential difference should be greater than or equal to 150 mv. | ||||
— | — | — |
3. Calibration of Burette :
a. Calibration of Burette for Karl Fisher – 10 ml :
Burette No. : _______________ Balance ID: _________________
Cal. For (ml) | Wt. of Beaker (g) | Wt. of Beaker + water (g) | Difference in wt. (g) | Difference in terms of ml | Burette reading displayed (ml) | Error in ml | Error in % | Limit | Remarks |
— | [A] | [B] | [C= B – A] | {C/0.9970} | — | — | — | — | — |
2 | ± 0.30 % | ||||||||
5 | ± 0.20 % | ||||||||
8 | ± 0.20 % | ||||||||
10 | ± 0.20 % |
Conclusion: Calibration of Burette for Karl fisher – 10 mlcomplies/Does not complies as per acceptance criteria.
b. Calibration of Burette for Non Aqueous (auto Titrator) – 20 ml :
Burette No. : _______________ Balance ID: ________________
Cal. For (ml) | Wt. of Beaker (g) | Wt. of Beaker + Water (g) | Difference in wt. (g) | Difference in terms of ml | Burette reading displayed (ml) | Error in ml | Error in % | Limit | Remarks |
— | [A] | [B] | [C= B – A] | {C/0.9970} | — | — | — | — | — |
2 | ± 0.30 % | ||||||||
5 | ± 0.20 % | ||||||||
8 | ± 0.20 % | ||||||||
10 | ± 0.20 % | ||||||||
20 | ± 0.15 % |
Conclusion: Calibration of Burette for Karl fisher – 20 mlcomplies/Does not complies as per acceptance criteria.
c. Calibration of Burette for Aqueous (auto Titrator) – 20 ml :
Burette No. : _______________ Balance ID: _________________
Cal. For (ml) | Wt. of Beaker (g) | Wt. of Beaker + Water (g) | Difference in wt. (g) | Difference in terms of ml | Burette reading displayed (ml) | Error in ml | Error in % | Limit | Remarks |
[A] | [B] | [C= B – A] | {C/0.9970} | ||||||
2 | ± 0.30 % | ||||||||
5 | ± 0.20 % | ||||||||
8 | ± 0.20 % | ||||||||
10 | ± 0.20 % | ||||||||
20 | ± 0.15 % |
Conclusion:
Calibration of Burette for Karl fisher – 20 mlcomplies/Does not complies as per acceptance criteria.
4. Performance Verification of double Pt. wire electrode (KF titration):
Electrode ID: _______________
Lot no. of Di-sodium Tartrate: ________________
Date | Weight of Di-sodium Tartrate | Water content (%) | Acceptance Criteria |
15.66 ± 3% (15.190 – 16.130) | |||
— | — |
Conclusion:
Performance Verification of double Pt. wire electrode (KF titration) complies/Does not complies as per acceptance criteria.
5. Standardization by Potentiometric :
Electrode ID: _______________
Lot no. of KHC8H4O4: ________________
Sample ID | Weight of KHC8H4O4 | Normality (N) | Limit |
1 | N ± 5 % (0.0950 N – 0.1050 N) | ||
2 | — | ||
3 | — | ||
— | % RSD | — | NMT 0.2 % |
Conclusion:
Standardization by potentiometric isComplies/does not complies as per acceptance
Criteria.
Remarks: Calibration Satisfactory / Not Satisfactory
— | Calibrated By | Checked By | Approved By |
Signature | |||
Date | |||
Name | |||
Department | Quality Control | Quality Control | Quality Control |