Calibration Protocol of Auto Titrator

Instrument Calibration Protocol of Auto Titrator (905 titrando)

INSTRUMENTAuto Titrator (905 titrando)
Refer SOP No. 
Model No.                     905 titrando
Instrument Code No.                
MakeMetrohm
FrequencyQuarterly “- 5 days”
Date of Calibration                
Next Due Date 

1. Calibration of pH glass electrode (Aqueous) :

[ Electrode ID :______________________ ]

DateBufferPotential valueSlope  pH of buffer 7.00Acceptance criteria
     Slope should between 96-102% Buffer 7.00 pH – should between 6.75-7.25
  
  

Conclusion: Calibration of pH glass electrode (Aqueous)Complies/Does not complies as per

                     Acceptance criteria.

2. Calibration of Non – aqueous electrode (LL Solvotrode) :

[Electrode ID: ______________________]

DateBufferPotential valuePotential   DifferenceAcceptance criteria
    Potential difference should be greater than or equal to 150 mv.
  

3. Calibration of Burette  :

a. Calibration of Burette for Karl Fisher – 10 ml :

Burette No. : _______________      Balance ID: _________________

Cal. For (ml)Wt. of Beaker (g)Wt. of Beaker + water (g)Difference in wt. (g)Difference in terms of mlBurette reading displayed (ml)Error in mlError in %LimitRemarks
[A][B][C= B – A]{C/0.9970}
2       ± 0.30 % 
5       ± 0.20 % 
8       ± 0.20 % 
10       ± 0.20 % 

Conclusion: Calibration of Burette for Karl fisher – 10 mlcomplies/Does not complies as per acceptance criteria.

b. Calibration of Burette for Non Aqueous (auto Titrator) – 20 ml :

                      Burette No. : _______________      Balance ID: ________________

Cal. For (ml)Wt. of Beaker (g)Wt. of Beaker + Water (g)Difference in wt. (g)Difference in terms of mlBurette reading displayed (ml)Error in mlError in %LimitRemarks
[A][B][C= B – A]{C/0.9970}
2       ± 0.30 % 
5       ± 0.20 % 
8       ± 0.20 % 
10       ± 0.20 % 
20       ± 0.15 % 

Conclusion:   Calibration of Burette for Karl fisher – 20 mlcomplies/Does not complies as per acceptance criteria.

c. Calibration of Burette for Aqueous (auto Titrator)  – 20 ml :

Burette No. : _______________      Balance ID: _________________

Cal. For (ml)Wt. of Beaker (g)Wt. of Beaker + Water (g)Difference in wt. (g)Difference in terms of mlBurette reading displayed (ml)Error in mlError in %LimitRemarks
 [A][B][C= B – A]{C/0.9970}     
2       ± 0.30 % 
5       ± 0.20 % 
8       ± 0.20 % 
10       ± 0.20 % 
20       ± 0.15 % 

Conclusion:

Calibration of Burette for Karl fisher – 20 mlcomplies/Does not complies as per acceptance criteria.

4. Performance Verification of double Pt. wire electrode (KF titration):

           Electrode ID: _______________

  Lot no. of Di-sodium Tartrate: ________________

DateWeight of Di-sodium TartrateWater content (%)Acceptance Criteria
   15.66 ± 3%                 (15.190 – 16.130)
  

Conclusion:

Performance Verification of double Pt. wire electrode (KF titration) complies/Does not complies as per acceptance criteria.

 5. Standardization by Potentiometric :

          Electrode ID: _______________

   Lot no. of KHC8H4O4: ________________

Sample IDWeight of KHC8H4O4Normality (N)Limit
1  N ± 5 % (0.0950 N – 0.1050 N)
2  
3  
% RSDNMT 0.2 %

Conclusion:

       Standardization by potentiometric isComplies/does not complies as per acceptance

       Criteria.

Remarks: Calibration Satisfactory / Not Satisfactory

      Calibrated ByChecked ByApproved By
Signature   
Date   
Name   
DepartmentQuality ControlQuality ControlQuality Control