Instrument Calibration Protocol of Temperature Control Module
INSTRUMENT: Temperature Control Module |
Refer SOP No.: |
Model No.: TC2 |
Instrument Code No.: |
Make: Waters |
Frequency: Half Yearly – 15 days |
Date of Calibration: |
Next Due Date: |
Calibration of Oven:
Keep the temperature sensor in the column oven & close the door.
Set the different temperature 25°C, 40°C, 50°C, 70°C, 100°C and observe the temperature on temperature sensor and LCD screen.
Calibrated Temperature Sensor ID: __________________
Sr. No. | Set Temperature | Observed Temperature on Calibrated Temperature Sensor | Observed Temperature on LCD Screen | Difference |
1 | 25°C | |||
2 | 40°C | |||
3 | 50°C | |||
4 | 70°C | |||
5 | 100°C |
Acceptance Criteria: Difference should not be more than ± 2 °C of set temperature between calibrated temperature sensor and on LCD screen.
Remark: The instrument is calibrated & qualified/ Out of calibration & not qualified for use.
Calibration : Scheduled / Not scheduled
(Reason : ___________________________________________________________ )
Calibrated By | Checked By | Approved By | |
Signature | |||
Date | |||
Name | |||
Department |