Calibration Protocol of Temperature Control Module

Instrument Calibration Protocol of Temperature Control Module

INSTRUMENT: Temperature Control Module
Refer SOP No.:  
Model No.: TC2               
Instrument Code No.:          
Make: Waters
Frequency: Half Yearly – 15 days
Date of Calibration:       
Next Due Date:

Calibration of Oven:

Keep the temperature sensor in the column oven & close the door.

Set the different temperature 25°C, 40°C, 50°C, 70°C, 100°C and observe the temperature on temperature sensor and LCD screen.

Calibrated Temperature Sensor ID: __________________

Sr. No.Set TemperatureObserved Temperature on Calibrated Temperature SensorObserved Temperature on LCD ScreenDifference
125°C   
240°C   
350°C   
470°C   
5100°C   

Acceptance Criteria: Difference should not be more than ± 2 °C of set temperature between calibrated temperature sensor and on LCD screen.

Remark: The instrument is calibrated & qualified/ Out of calibration & not qualified for use.

Calibration : Scheduled / Not scheduled

(Reason : _____________­­­­­­­­­­______________________________________________ )

   Calibrated By   Checked By   Approved By
Signature   
Date   
Name   
Department   