Calibration Protocol of Thermo gravimetric Analyzer (TGA)

Instrument Calibration Protocol of Thermo gravimetric Analyzer (TGA)

Instrument ID. No. 
Reference SOP No.
Model No.                        Q500
MakeTA Instruments
Frequency:Quarterly – 5 days.
Date of Calibration              
Next Due Date

A. Traceable Mass (Weight) Calibration:

Sr. No.DescriptionExpected ValuesActual ValuesResult
1Mass 100 mg calibration99.90 to 100.10 mg   Pass/ Fail
2  Mass 1 g calibration0.999 to 1.001 g   Pass/ Fail

Done By :                                                                   Checked By :

B. Traceable Temperature Calibration:

Sr. No.DescriptionValues ExpectedValues ActualResult
1Alumel curie temperature148.9 to 159.5ºC   Pass/ Fail
2  Nickel curie temperature353.0 to 363.6ºC  Pass/ Fail

Done By :                                                                   Checked By :

Remarks: The instrument is calibrated & qualified / Out of calibration & not qualified for use.

Calibrated ByChecked ByApproved By
Signature
Date
Name
DepartmentQuality ControlQuality ControlQuality Control