Calibration Protocol of Vacuum Oven

Instrument Calibration Protocol of Vacuum Oven

InstrumentVacuum Oven
Instrument Code Number 
Refer Operating SOP No. 
Model No. 
MakeBINDER
Date of Calibration 
Next Due Date 

1. Operating Range:

Calibration Certificate No.:

Sr. No.Calibrated Temperature Sensor ID No.Calibration Due On
1  
2  
3  
4  
5  

a. Set Temperature: 50°C

Sr. No.LocationTime IntervalTemperature Observed Instrument Display (ºC)Temperature Observed Temperature sensor (ºC)Difference
1Top – Left1 Hour   
2 Hour   
2Top – Right1 Hour   
2 Hour   
3Middle1 Hour   
2 Hour   
4Bottom – Left1 Hour   
2 Hour   
5Bottom – Right1 Hour   
2 Hour   

Acceptance Criteria: ±1°C of the Set Temperature.

Done by:                                                                    Checked by:

b. Set Temperature : 105°C

Sr. No.LocationTime IntervalTemperature Observed Instrument Display (ºC)Temperature Observed Temperature sensor (ºC)Difference
1Top – Left1 Hour   
2 Hour   
2Top – Right1 Hour   
2 Hour   
3Middle1 Hour   
2 Hour   
4Bottom – Left1 Hour   
2 Hour   
5Bottom – Right1 Hour   
2 Hour   

Acceptance Criteria: ±1°C of the Set Temperature.

Done by:                                                                    Checked by:

c. Set Temperature : 200°C

Sr. No.LocationTime IntervalTemperature Observed Instrument   Display (ºC)Temperature Observed Temperature sensor (ºC)Difference
1Top – Left1 Hour   
2 Hour   
2Top – Right1 Hour   
2 Hour   
3Middle1 Hour   
2 Hour   
4Bottom – Left1 Hour   
2 Hour   
5Bottom – Right1 Hour   
2 Hour   

Acceptance Criteria: ± 1°C of the Set Temperature.

Done by:                                                                    Checked by:

Remarks: Verification Satisfactory / Not Satisfactory.

Calibrated ByChecked ByApproved By
Signature   
Date   
Name   
DepartmentQuality ControlQuality ControlQuality Control