Calibration Protocol of Spectrofluorometer

Instrument Calibration Protocol of Spectrofluorometer

INSTRUMENTSpectrofluorometer
Instrument Code No. 
Reference SOP No. 
Model No.FP-8300
MakeJASCO
Date of Calibration 
Next Due Date 
FrequencyQuarterly -5 days

1. Wavelength Accuracy: For Excitation Monochromator :

Sr. No.MeasuredMean DifferenceCriteriaResult
1  253.7 ± 1.5 nm (252.2 nm to 255.2 nm)Pass □              Fail □
2  
3  
1  365.0 ± 1.5 nm (363.5 nm to 366.5 nm)Pass □              Fail □
2  
3  
1  435.8 ± 1.5 nm (434.3 nm to 437.3 nm)Pass □              Fail □
2  
3  
1  546.1 ± 1.5 nm (544.6 nm to 547.6 nm)Pass □              Fail □
2  
3  

Done By :                                                                   Checked By :

For Emission Monochromator:

Sr. No.MeasuredMean DifferenceCriteriaResult
1  253.7 ± 1.5 nm (252.2 nm to 255.2 nm)Pass □              Fail □
2  
3  
1  365.0 ± 1.5 nm (363.5 nm to 366.5 nm)Pass □              Fail □
2  
3  
1  435.8 ± 1.5 nm (434.3 nm to 437.3 nm)Pass □              Fail □
2  
3  
1  546.1 ± 1.5 nm (544.6 nm to 547.6 nm)Pass □              Fail □
2  
3  

Done By :                                                                   Checked By :

2. Wavelength Repeatability:

For Excitation Monochromator:

Sr. No.MeasuredMean DifferenceCriteriaResult
1  546.1 ± 1.0 nm (545.1 nm to 547.1 nm)Pass □              Fail
2  
3  

Done By :                                                                   Checked By :

For Emission Monochromator :

Sr. No.MeasuredMean DifferenceCriteriaResult
1  546.1 ± 1.0 nm (545.1 nm to 547.1 nm)Pass □              Fail □
2  
3  

Done By :                                                                   Checked By :

3. Sensitivity:

Peak height of Water Raman Peak (Height of Peak – Height of Baseline) =

      Noise Level (Maximum noise level – Minimum noise level) =

Ratio = Peak height of Water Raman Peak/ Noise Level

Ratio =

Signal to noise ratio

Peak height of Water Raman PeakNoise LevelRatioCriteriaResult
   NLT 170Pass □          Fail □

Done By :                                                                Checked By :

  Remarks: The instrument is calibrated & qualified / Out of calibration & not qualified for use.

Calibrated ByChecked ByApproved By
Signature   
Date   
Name   
Department   Quality ControlQuality ControlQuality Control