Instrument Calibration Protocol for FT-IR (SHIMADZU)
Instrument | : | FT-IR |
Instrument Code Number | : | D/QCD/FTR01 |
Refer Operating SOP No. | : | DQC-SOP-0066 |
Model No. | : | FTIR-8400 |
Date of Calibration | : | |
Make | : | SHIMADZU |
Next Due Date | : |
1. Perform auto adjustment:
Observation: Auto adjustment found satisfactory / not satisfactory.
2. Verification of Wave number Scale :
Sr. No. | Peak to be observe at Wave numbers | Measured Wave number | Error | Limits in wave numbers (cm-1) | Remark |
1. | 3060.0 cm-1 | 3059.0 – 3061.0 cm-1 | |||
2. | 2849.5 cm-1 | 2848.5 – 2850.5 cm-1 | |||
3. | 1942.9 cm-1 | 1941.9 – 1943.9 cm-1 | |||
4. | 1601.2 cm-1 | 1600.2 – 1602.2 cm-1 | |||
5. | 1583.0 cm-1 | 1582.0 – 1584.0 cm-1 | |||
6. | 1154.5 cm-1 | 1153.5 – 1155.5 cm-1 | |||
7. | 1028.3 cm-1 | 1027.3 – 1029.3 cm-1 |
3. Resolution performance of Instrument :
Sr. No. | % Transmittance difference between | Observed Results | Limits | Remarks |
1. | 2849.5 cm-1 & 2870 cm-1 | NLT 0.33 | ||
2. | 1583 cm-1 & 1589 cm-1 | NLT 0.08 |
4. Verification of Absorbance Repeatability :
Sr.No. | Absorbance to be Observe at Wave numbers (cm-1) | ABS | ABS | Error | Tolerance |
1. | 2849.5 | ± 0.03 cm-1 | |||
2. | 1601.2 | ± 0.05 cm-1 | |||
3. | 1028.3 | ± 0.03 cm-1 |
5. Verification of Wave number Repeatability :
Corresponding Wave number (cm-1) | (I ) Observed Wave number (cm-1) | (II) Observed Wave number (cm-1) | Difference in Wave numbers (cm-1) | Limits (cm-1) | Remarks |
2849.5 | ± 5 cm-1 | ||||
1601.2 | ± 1 cm-1 | ||||
1028.3 | ± 1 cm-1 |
6. Verification of Transmittance Repeatability :
Corresponding Wave number (cm-1 ) | % T (I) | % T (II) | Difference in % T | Limits (%) | Remarks |
2849.5 | ± 0.5 cm-1 | ||||
1601.2 | ± 0.5 cm-1 | ||||
1028.3 | ± 0.5 cm-1 |
Remarks: Calibration satisfactory / Not satisfactory
Calibrated By
Signature/ Date
Checked By
Signature/ Date
Approved By
Signature/ Date