Ion Chromatography Calibration Protocol
| Instrument : Ion Chromatography |
| Instrument Code Number : |
| Refer Operating SOP No. : |
| Model No : |
| Make: |
| Date of Calibration : |
| Next due date: |
1. Detector Linearity:
Nitrate Standard Solution 1000mg/lNO3
Batch / Lot No. : ________________ Expiration date: ___________
Observation Table:
| Sr. No. | Concentration (ppm) | Height (µS) |
| Concentration Vs Height | ||
| Observed Correlation coefficient | ||
| Limit | ≥ 0.999 | Should be Linear. |
2. Noise Drift:
Sample: Inject 25 µL of Milli Q Water once.
Observation Table:
Data for Detector Noise
| Segment No. | Noise, nS |
| Average, nS | |
| Limit, nS | 2.0 |
| Result |
Data for Detector Drift
| 20 minute, Drift, nS | Drift, nS/hr | Limit, nS/hr | Result |
| 20 |
3. Injector Precision:
Note: Run the single injection of DI (De Ionized) water with full loop capacity before standard injection.
Nitrate Standard Solution 1000mg/lNO3
Batch / Lot No.: ________________,
Expiration date: ____________
Observation Table:
| Name | Area (µS * min) |
| Average | |
| Std dev. | |
| % RSD | |
| Limit | 1.0 % |
| Result |
4. Injector Linearity:
Nitrate Standard Solution 1000mg/lNO3
Batch / Lot No. : ________________, Expiration Date: ____________
Observation Table:
| Sr. No. | Injection volume (µl) | Area (µS * min) |
| Injection volume Vs Area | ||
| Correlation coefficient | ||
| Limit | ≥ 0.999 | Should be Linear. |
5. Injector Carry Over Check:
Observation Table:
| Name | Retention time (min) | Area (µS * min) |
| Carry over | – | |
| Limit | – | 0.1 % |
| Result | – |
6. Gradient Pump Accuracy Test:
Inject 25 µl of Milli Q water.
Observation Table:
Step Accuracy – Valves A, B
| Theoretical value (%) | Conductivity (µS) | Calculated value (%) | Deviation (%) | Limit (±, %) | Result |
| N/A | – | ||||
| 2.0 | |||||
| 2.0 | |||||
| 2.0 | |||||
| 2.0 |
Step Accuracy – Valves C, D
| Theoretical value (%) | Conductivity (µS) | Calculated value (%) | Deviation (%) | Limit (±, %) | Result |
| N/A | – | ||||
| 2.0 | |||||
| 2.0 | |||||
| 2.0 | |||||
| 2.0 |
Step Noise – Valves A, B
| Step (%) | Noise (µS) | Calculated Noise (%FS) | Limit ( %) | Result |
| NMT 0.20% | ||||
| NMT 0.20 % | ||||
| NMT 0.20 % | ||||
| NMT 0.20 % | ||||
| NMT 0.20 % |
Step Noise – Valves C, D
| Step (%) | Noise (µS) | Calculated Noise (%FS) | Limit ( %) | Result |
| NMT 0.20 % | ||||
| NMT 0.20 % | ||||
| NMT 0.20 % | ||||
| NMT 0.20 % | ||||
| NMT 0.20 % |
7. Flow rate Accuracy and Precision Test:
Balance ID: _________________
Weight of 25 ml measuring cylinder: _________________ g
Temperature: _________________
Observation Table:
| Segment | Weight of measured eluent in measuring cylinder in ‘g’ | Calculated eluent i.e. Calculated Flow rate in ml/min. | Deviation from 1.00 ml/min. (Limit: NMT ± 0.050) |
| Average Flow | |||
| Std Dev. | |||
| % RSD (Limit: NMT 2.0 %) |
REMARKS: Calibration satisfactory /Not satisfactory
Prepared By
Signature/ Date
Checked By
Signature/ Date
Approved By
Signature/ Date

