Calibration Protocol for Atomic Absorption Spectrometer (AAS)

Instrument Calibration Protocol for Atomic Absorption Spectrometer (AAS)

InstrumentAtomic Absorption Spectrometer
Instrument ID. No. 
Ref. SOP No. 
Model No.                         
Calibration frequencyHalf Yearly (-15 days).
Date of calibration               
Next due date 

Note:

  • Prior to start calibration activity, required glassware shall be rinsed with the respective acid diluents as per standard bottle label. For example, use 1% HNO3 in case of calibration performing with Manganese & Nickel use 1% HCl in case of calibration performing with Chromium standard.
  •   The blank shall be prepared as similar to the rinse volumetric flask with the acid diluents. For 

  Example, use 1% HNO3 in case of calibration performing with Manganese & Nickel, and use

  1% HCl in case of calibration performing with Chromium standard.

  • The graphite furnace auto sampler wash vessel must be filled with suitable wash diluents. For example: Use 1% HNO3 for Manganese standard solution as a wash diluent.
  • Prior to calibration, graphite shall be checked visually and if any carbon deposition observed it shall be replaced with the new one.

1. Air-Acetylene flame test using Nickel Standard:

     Nickel Standard Batch Number:

           Validity:

           Preparation of 1% HNO3:

Preparation of Nickel Standard (10 ppm):

Pipetted out _______ (1.0 mL) of 1000 mg/L Nickel standard solution in 100 mL volumetric    

flask and diluted it up to the mark with water, Shaked well.

Sr. No.Name of StandardNo. of ResamplesLimitResult% RSD (NMT 1.0%)
1Nickel (10 ppm)10Absorbance (NLT 0.50)  

Remark: The Instrument complies / does not comply Air-Acetylene flame test by using Nickel Standard.

  Analyzed By:                                                                        Checked By:

  Date:                                                                                      Date:

2. Nitrous Oxide-Acetylene flame test using Chromium standard:

Chromium Standard Batch Number:

Validity:

Preparation of 1% HCl:

Preparation of Chromium Standard (3ppm):

1) Pipetted out _______ (1.0 mL) of 1000 mg/L Chromium standard solution in 10 mL volumetric flask and diluted it upto the mark with water, shaked well.

2) Further pipetted out _______ (3.0 mL) of above solution in 100 mL volumetric flask and diluted it upto the mark with water, shaked well.

Sr. NoName of StandardNo. of ResamplesLimitResult% RSD (NMT 1.5%)
1Chromium (3 ppm)10Absorbance (NLT 0.16)  

Remark: The Instrument complies / does not comply Nitrous Oxide-Acetylene flame test using Chromium standard.

  Analyzed By:                                                                                   Checked By:

  Date:                                                                                                Date:

3. Furnace test using Manganese standard:

Note: Graphite visually checked is found Ok / Not Ok

Manganese Standard Batch Number:                        

Validity:        

Preparation of 1% HNO3:

Preparation of Manganese Standard (20ppb):

1) Pipetted out ________ (1.0 mL) of 1000 mg/L Manganese standard in 10 mL volumetric flask and diluted it upto the mark with water, shaked well.

2) further, pipetted out _______ (1.0 mL) of above solution in 100 mL volumetric flask and diluted it upto the mark with water, shaked well.

3)  further, pipetted out _______ (2.0 mL) of above solution in 100 mL volumetric flask and diluted it upto the mark with water, shaked well.

Sr. NoName of StandardNo. of ResamplesLimitResult% RSD (NMT 5.0%)
1Manganese (20 ppb)5Absorbance (NLT 0.80)                

Remark: The Instrument complies / does not comply furnace test using Manganese standard.

Analyzed By:                                                                      Checked By:

Date:                                                                                  Date:

Conclusion: Calibration found satisfactory/not satisfactory.

Calibrated By (Sign/ Date)

Checked By (Sign/ Date)