Calibration Protocol for UV-Visible Spectrophotometer

Instrument Calibration Protocol for UV-Visible Spectrophotometer

INSTRUMENTUV-Visible Spectrophotometer
Instrument ID. No.
Ref. SOP No.
Model No.                        Perkin
Calibration frequency 
Date of calibration               
Next due date 

1. Control of Absorbance :

  Balance ID : _________________

  Potassium Dichromate Lot No. / Batch No. : ___________________

 Potassium Dichromate Expiry / Valid Up to : ___________________

For 235 nm, 257 nm, 313 nm and 350 nm :

Weight of Potassium Dichromate: ___________ g.

For 430 nm :

Weight of Potassium Dichromate: ___________ g.

Wavelength (nm)AbsorbanceA (1%, 1 cm)STD A (1 %, 1 cm)LimitRemarks
235  124.5122.9 to 126.2 
257  144.0142.8 to 146.2 
313  48.647.0 to 50.3 
350  106.6105.6 to 109.0 
430  15.915.7 to 16.1 

Where,

A (%, 1cm) = [A / (C x D)]

A – Observed Absorbance

C – Concentration of Potassium dichromate solution in % as w/v.

D – Path Length in cm.

2. Resolution :

Spectroscopic Hexane Lot No. / Batch No. : ___________________

Spectroscopic Hexane Expiry / Valid Up to : ___________________

Spectroscopic Toluene Lot No. / Batch No. : ___________________

Spectroscopic Toluene Expiry / Valid Up to : ___________________

Observation :   

Absorbance at about 269 nm : _________

Absorbance at about 266 nm : _________

Ratio : = Maximum Absorbance/ Minimum Absorbance =

Limit: Not Less than 1.5.

3. Limit of Stray Light :

Balance ID : _________________

Potassium Chloride Lot No. / Batch No. : ___________________

Potassium Chloride Expiry / Valid Up to : ___________________

Weight of Potassium Chloride  : ___________ g.

Observation :

Absorbance at 200 nm : _________

Limit   : Not Less than 2.0.            

4. Baseline Flatness :

Observation : Observed Baseline Flatness : _________

Limit   :  Baseline Flatness shall be within ± 0.002 Abs.

5. Second derivative Spectrophotometer : Resolution Power :

 Spectroscopic Toluene Lot No. / Batch No.  : ___________________

Spectroscopic Toluene Expiry / Valid Up to : ___________________

Methanol AR Grade Lot No. / Batch No.      : ___________________

Methanol AR Grade Expiry / Valid Up to     : ___________________

 Observation :

Ratio : A/ B =

Limit   : Not less than 0.2.

6. Control of Wavelength:         

Balance ID: _________________

Holmium Oxide AR Grade Lot No. / Batch No. : ___________________

Holmium Oxide AR Grade Expiry / Valid Up to : ___________________

Weight of Holmium Oxide AR: ___________ g.          

Wavelength (nm)Observed Wavelength (nm)Maximum ToleranceRemarks
241.15 240.15 to 242.15 
287.15 286.15 to 288.15 
361.50 360.50 to 362.50 
536.30 533.30 to 539.30 

7. Control of Wavelength (In House Method) :

Balance ID : _________________

Caffeine AR Grade Lot No. / Batch No. : ___________________

Caffeine AR Grade Expiry / Valid Up to : ___________________

Weight of Caffeine AR   : ___________ mg.

Wavelength (nm)Observed Wavelength (nm)Maximum ToleranceRemarks
205 204.0 – 206.0 
273 272.0 – 274.0 

Remarks: Calibration Satisfactory / Not Satisfactory

Calibrated By (Sign/ Date)

Checked By (Sign/ Date)