Calibration protocol for Friability Test Apparatus

Instrument calibration protocol for Friability Test Apparatus

InstrumentFriability Test Apparatus
Frequency  :Monthly
Refer operation SOP No.      
Instrument Code No.   
MakeElectrolab
Standard stop watch code no.: 
Calibration of stopwatch valid up to 
Date of Calibration 
Next due date : 

1. Calibration of Rotation :

Switch “ON” the Mains. Switch “ON” the Instrument.

Set the Rotation to 25.

Start the Instrument and calibrated stopwatch at a same time.

Count the Rotation per 60 Second.

Repeat the procedure for two more times.

Repeat the Step 2 to 5 for Rotation 50, 75 and 100 (120, 180 and 240 seconds, respectively).

RPMRun IRun IIRun IIILimit
25   25 ± 1 Rotation
50   50 ± 2 Rotation
75   75 ± 3 Rotation
100   100 ± 4 Rotation

            Done by :                                                                    Checked by :

2. Calibration of Time :  

1. Switch “ON” the Mains. Switch “ON” the Instrument.

2. Set the Time to 60 Second.

3. Start the Instrument and calibrated stopwatch at a same time.

4. Note the time in the stopwatch at the moment the drums stops.

5. Repeat the procedure for two more times.

6. Repeat the Step 2 to 5 for time 120, 180 and 240 second.

Time (Sec)Run IRun IIRun IIILimit
60   60 ± 2.5 Second
120   120 ± 5 Second
180   180 ± 7.5 Second
240   240 ± 10 Second

Done by :                                                                                              Checked by :

3. Calibration of LCD Display :

Set the Rotation 100.

Count the Rotation. When 100 rotations are completed, check the display.

Fill the Calibration Check Report and check for meeting the Acceptance Criteria.

Repeat the procedure for three more times and average the LCD display for Rotation.

Set Rotation: 100.

Observations of Rotation Displayed on LCD1st2nd3rd4thAverage
Reading     

Limits: 100 ± 1 Rotations

Done by:                                                                    Checked by:

Remarks: Calibration Satisfactory / Not Satisfactory

Calibrated By (Sign/ Date)

Checked By (Sign/ Date)