Mechanical Calibration Protocol for Dissolution Apparatus
INSTRUMENT | Dissolution Test Apparatus |
Refer SOP No. | DQC-SOP-0612 |
Model No. | EDT 14 Lx |
Instrument ID. No. | |
Make | ELECTROLAB |
Frequency | Quarterly “- 5 Days” |
Date of Calibration | |
Next Due Date |
1. Visual Inspection of Vessel, Basket and Paddle :
Vessel Examination:
Vessels shall be checked visually for the chipping, Plastic/glass vessels shall be checked for the scratching and Baskets and Paddles for the distortion.
Conclusion: found satisfactory / Not satisfactory.
Basket Examination:
Each Basket shall be visually examined for defects such as rusting, the edge of the wire mesh sticking out beyond the Basket, Clogged Mesh Holes and smashed in mesh sides.
Conclusion: found satisfactory / Not satisfactory.
Paddle Examination:
Each Paddle shall be visually examined for defects such as rusting and loose pieces of coating sticking out from the Paddles.
Conclusion: found satisfactory / Not satisfactory.
2. Calibration of Rotation per Minute (RPM) :
[Calibrated Digital Tachometer ID: _________________________] [Valid up to: _______________]Sr. No. | Set RPM | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 | 10 | 11 | 12 | 13 | 14 |
1 | 50 | ||||||||||||||
2 | 100 | ||||||||||||||
3 | 150 | ||||||||||||||
4 | 200 | ||||||||||||||
5 | 300 |
Acceptance Criteria: ± 1 Rotation per Minute (RPM).
Done by: Checked by:
3. Calibration of Temperature :
A. Water Bath Temperature Calibration:
[Calibrated Thermometer ID: _______________] [Valid up to: _______________]Sr. No. | Set Temperature | Displayed Temperature (°C) | Actual temp. in Water bath (Calibrated Thermometer) (°C) | Difference | Remarks |
1 | 35.0 °C | ||||
2 | 37.0 °C | ||||
3 | 39.0 °C |
Acceptance Criteria: The Difference between Calibrated Thermometer temperature and displayed temperature should Not More Than ± 0.5 °C.
Done by: Checked by:
B. Calibration of Jar Temperature :
[Calibrated Thermometer ID: _______________] [Valid up to: _______________]Place the vessels containing 500mL of water (at room temperature).
Set the temperature to 37.0°C.
Allow the instrument to stabilize. Verify the temperature in each fourteen vessel.
Set Temperature: 37.0 °C
Jar No. | Displayed Temperature (°C) | Actual Temperature in Jar (Calibrated Thermometer) (°C) | Difference | Remarks |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
7 | ||||
8 | ||||
9 | ||||
10 | ||||
11 | ||||
12 | ||||
13 | ||||
14 |
Acceptance Criteria: The Difference between Calibrated Thermometer temperature and displayed temperature value Not more than ± 0.5 °C.
Done by: Checked by:
4. Calibration of Time :
[Calibrated Stop Watch ID: _________________] [Valid up to: _______________]Sr. No. | Set Time (Minutes) | Displayed Time (Minutes) | Actual Time in Stop Watch (Minutes) | Difference (Minutes) | Remarks |
1 | 10 | ||||
2 | 20 | ||||
3 | 30 |
Acceptance Criteria: The Difference between Displayed time and stop watch should not be more than ± 5 Seconds.
Done by: Checked by:
5. Calibration of Wobbling, Centering and D1 and D2 Distance :
Calibrated Wobbling Gauge ID: ______________, Valid up to : __________
Calibrated Depth Gauge ID: __________________, Valid up to : __________
Sr. No. | Limit of Wobbling Paddle | Limit of Wobbling Basket | Limit of D1 Distance | Limit of D2 Distance |
— | NMT 0.5 mm | NMT 1.0 mm | 25.0 ± 2.0 mm | 25.0 ± 2.0 mm |
1 | ||||
2 | ||||
3 | ||||
4 | ||||
5 | ||||
6 | ||||
7 | ||||
8 | ||||
9 | ||||
10 | ||||
11 | ||||
12 | ||||
13 | ||||
14 |
Where,
D1 = Distance between the bottom edge of the basket to the lowest inner surface of the vessel.
D2 = Distance between the bottom edge of the blade to the lowest inner surface of the vessel.
Centering :
Calculate the centering using below formula;
C = Sqrt [(x1-x2)2 + (y1-y2)2] / 2
Where,
x1, x2, y1 and y2 are the distance from the inner vessel wall at four location equally spaced to shaft by vernier caliper.
Observation:
Calibrated Centering kit ID: __________________, Valid up to: ___________
Calibrated Vernier caliper ID :________________, Valid up to : ___________
Limit for Centering, c : NMT 2.0mm
Vessel | x1 mm | x2 mm | y1 mm | y2 mm | x1-x2 | y1-y2 | Centering, c (mm) |
1 | |||||||
2 | |||||||
3 | |||||||
4 | |||||||
5 | |||||||
6 | |||||||
7 | |||||||
8 | |||||||
9 | |||||||
10 | |||||||
11 | |||||||
12 | |||||||
13 | |||||||
14 |
Calculation: (Vessel -1):
Calculate the same for other vessel.
Done by: Checked by:
6. Leveling Check:
By Digital protractor
[Calibrated Digital protractor ID: ____________________, Valid up to: ___________]Verify the Bench top levelness and Base Plate (Vessel support plate) levelness. Measure the inclination on the each side (i.e. On four sides front, rear, left and right) for Bench Top and for Base plate. If leveling is not satisfactory for Base plate, adjust the level by rotating the adjusting screws on the feet of the assembly.
Observation table:
— | Front | Rear | Left | Right | Limit |
Bench Top Inclination | NMT 1° | ||||
Base Plate Inclination | NMT 0.5° |
Conclusion: Leveling found satisfactory / notsatisfactory limit.
Done by: Checked by:
7. Alignment verification by Digital Protractor :
[Calibrated Digital protractor ID: ____________________, Valid up to: ___________]Shaft Verticality:
Measure the verticality of paddle shaft and basket shaft in the dissolution assembly.
Measurement of the verticality of the stirring element.
Vessel Verticality
Measure the verticality of vessel in the dissolution assembly. Take the measurement on the vessel inner surface.
Observation table:
Shaft No. # | Paddle Shaft Limit (90°±0.5°) | Basket Shaft Limit (90°±0.5°) | Vessel Limit (90°±0.5°) |
1. | |||
2. | |||
3. | |||
4. | |||
5. | |||
6. | |||
7. | |||
8. | |||
9. | |||
10. | |||
11. | |||
12. | |||
13. | |||
14. |
Conclusion: Shaft and Vessel verticality found satisfactory / notsatisfactory limit.
Done by: Checked by:
Remark: The instrument is calibrated & qualified/ Out of calibration & not qualified for use.
Calibration: Scheduled / Not scheduled
(Reason: ___________________________________________________)
Remarks: Calibration Satisfactory / Not Satisfactory
Calibrated By (Sign/ Date)
Checked By (Sign/ Date)