Instrument Calibration Protocol for Gas Chromatography – Shimadzu
INSTRUMENT | Gas chromatography (Shimadzu) |
Instrument code no. | |
Ref. SOP No. | |
Make / Model No. | |
Calibration Frequency | Half Yearly “-15 days”. |
Date of Calibration | |
Next Due Date |
1. Flow rate calibration:
Flow Meter ID: ___________________]
Set flow rate (mL / min) | Observed flow rate (mL /min) – 1 | Observed flow rate (mL /min) – 2 | Observed flow rate (mL /min) – 3 | Mean flow rate (mL / min) | Acceptance criteria (mL / min) |
5 mL | 4.75 to 5.25 | ||||
20 mL | 19.0 to 21.0 | ||||
30 mL | 28.5 to 31.5 | ||||
40 mL | 38.0 to 42.0 | ||||
50 mL | 47.5 to 52.5 |
Done By: Checked By:
2. Verification of temperature:
a. Column oven:
Instrument used for Calibration:
Instrument No. / Code No.
Calibration Date
Next Due Date of Calibration
Sr. No. | Set Temperature | Temperature observed on display | Temperature observed on Calibrator | Limits |
30 °C | 30 °C ± 2 °C | |||
50 °C | 50 °C ± 2 °C | |||
100 °C | 100 °C ± 2 °C | |||
150 °C | 150 °C ± 2 °C | |||
200 °C | 200 °C ± 2 °C | |||
250 °C | 250 °C ± 2 °C | |||
350° C | 350 °C ± 2 °C |
b. Injector:
Sr. No. | Set Temperature Injector A | Set Temperature Injector B | Temperature observed on display Injector A | Temperature observed on display Injector B | Limits |
1. | 50 °C | 50 °C | 50 °C ± 2 °C | ||
2. | 100 °C | 100 °C | 100 °C ± 2 °C | ||
3. | 150 °C | 150 °C | 150 °C ± 2 °C | ||
4. | 200 °C | 200 °C | 200 °C ± 2 °C | ||
5. | 250 °C | 250 °C | 250 °C ± 2 °C | ||
6. | 300 °C | 300 °C | 300 °C ± 2 °C | ||
7. | 350° C | 350° C | 350 °C ± 2 °C |
c. Detector:
Sr. No. | Set Temperature FID A | Set Temperature TCD B | Temperature observed on display FID A | Temperature observed on display TCD B | Limits |
1. | 50 °C | 50 °C | 50 °C ± 2 °C | ||
2. | 100 °C | 100 °C | 100 °C ± 2 °C | ||
3. | 150 °C | 150 °C | 150 °C ± 2 °C | ||
4. | 200 °C | 200 °C | 200 °C ± 2 °C | ||
5. | 250 °C | 250 °C | 250 °C ± 2 °C | ||
6. | 300 °C | 300 °C | 300 °C ± 2 °C | ||
7. | 350 °C | 350 °C | 350 °C ± 2 °C |
Done By: Checked By:
3. Auto Injector and Detector Calibration :
a. Calibration of FID by Linearity Measurement. (By Using Capillary column):
Solvent Name | Batch No. | Calibration Standard No. | Purity in % | Use Before |
Tetrahydrofuran | ||||
Isopropyl alcohol | ||||
Dimethylformamide |
(a) THF Internal standard (IS) solution preparation: Take ______ (1.0 gm) of ____________________ Tetrahydrofuran in a ________ (50ml) volumetric flask containing 10ml of DMF and dilute up to mark with __________DMF.
(b) Sample preparation
Stock solution: Weigh accurately _______( 2.0 g) of Isopropyl alcohol standard in a ______(50 ml) volumetric flask containing about 10 ml of DMF, dilute up to mark with ________DMF.
Further dilute this to prepare the different concentration as below,
- 800 ppm: Take _______(1.0 ml of stock solution) and _______(2.0 ml) of THF (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________DMF.
- 1600 ppm: Take _______(2.0 ml of stock solution) and _______(2.0 ml) of THF (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________DMF.
- 2400 ppm: Take _______(3.0 ml of stock solution) and _______(2.0 ml) of THF (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________DMF.
- 3200 ppm: Take _______(4.0 ml of stock solution) and _______(2.0 ml) of THF (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________DMF.
- 4000 ppm: Take _______(5.0 ml of stock solution) and _______(2.0 ml) of THF (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________DMF.
(c) Chromatographic condition :
Parameter | Test Conditions | Applied Conditions |
Column No. | ||
Column | AT-5 (30m x 0.53mm x 5.0µ) or Equivalent | |
Oven temperature | 100°C-2 min.- @ 30°C/min. 250°-1min. | |
Detector | Flame Ionization detector (FID) | |
Detector Temperature | 270°C | |
Head pressure | 30 kpa | |
Injection volume | 1.0 µl | |
Split ratio | 1:40 | |
Injector temperature | 200°C | |
Run time | 8 min. |
(d) Calibration procedure:
- Set up the instrument as per the conditions.
- Inject the samples three times from each solution and calculate the ratio of IPA and THF. Record the observation in table below.
- Calculate the %RSD of mean ratio. It should not be more than 10.0%.
- Plot a linearity curve of concentrations Vs corresponding mean area ratio, using least square method. Calculate the correlation coefficient (r) and record the observations in a table below.
(e) Observation Table
Concentration of IPA | Area of IPA X | Area of THF Y | Area Ratio X/Y | Mean Area Ratio and % RSD |
800 ppm | ||||
— | ||||
— | ||||
1600 ppm | ||||
— | ||||
— | ||||
2400 ppm | ||||
— | ||||
— | ||||
3200 ppm | ||||
— | ||||
— | ||||
4000 ppm | ||||
— | ||||
— |
Correlation Coefficient _____________ (Limit: NLT 0.990)
(f) Result: Complies / does not comply
Done By: Checked By:
b. Calibration of FID by Linearity Measurement. (By Using Pack column):
Solvent Name | Batch No. | Calibration Standard No. | Purity in % | Use Before |
Ethyl alcohol | ||||
Isopropyl alcohol |
Water | Ultra Purified (Milli-Q) |
(a) IPA Internal standard (IS) solution preparation: Weigh accurately _____________ (5.0 g) of Isopropyl alcohol standard in _______ (50 ml) volumetric flask containing 10 ml of water, dilute up to mark with ___________ water.
(b) Sample preparation
Stock solution: Weigh accurately _____________ (5.0 g) of ethyl alcohol standard in _______ (50 ml) volumetric flask containing 10 ml of water, dilute up to mark with ___________ water.
Further dilute this to prepare the different concentration as below,
- 2000 ppm: Take _______(1.0 ml) of stock solution and _______(5.0 ml) of IPA (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________water.
- 4000 ppm: Take _______(2.0 ml) of stock solution and _______(5.0 ml) of IPA (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________water.
- 6000 ppm: Take _______(3.0 ml) of stock solution and _______(5.0 ml) of IPA (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________water.
- 8000 ppm: Take _______(4.0 ml) of stock solution and _______(5.0 ml) of IPA (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________water.
- 10000 ppm: Take _______(5.0 ml) of stock solution and _______(5.0 ml) of IPA (IS) in ______(50 ml) volumetric flask, dilute up to mark with ________water.
(c) Chromatographic condition :
Parameter | Test Conditions | Applied Conditions |
Column No. | ||
Column | Porapak Q – 80 / 100#, 2M, S.S. Column. | |
Carrier gas | Nitrogen | |
Flow rate | 50 ml/min | |
Detector | FID | |
Column temperature | 180°C | |
Injector temperature | 250°C | |
Detector temperature | 250°C | |
Injection volume | 1.0 µl | |
Run Time | 15.0 min |
(d) Calibration procedure:
- Set up the instrument as per the conditions.
- Inject the samples three times from each solution and calculate the ratio of
Ethyl alcohol and IPA. Record the observation in table below.
- Calculate the %RSD of area ratio at each level. It should not be more than 10.0 %.
- Plot a linearity curve of concentrations Vs corresponding mean area ratio, using least square method. Calculate the correlation coefficient ® and record the observations in given table.
(e) Observation Table:
Concentration of Ethyl alcohol | Area of Ethyl alcohol X | Area of IPA Y | Area Ratio X/Y | Mean Area Ratio and % RSD |
2000 ppm | ||||
— | — | |||
— | — | |||
4000 ppm | ||||
— | — | |||
— | — | |||
6000 ppm | ||||
— | — | |||
— | — | |||
8000 ppm | ||||
— | ||||
— | ||||
10000 ppm | ||||
— | — | |||
— | — |
Correlation Coefficient: _____________ (Limit: NLT 0.990)
(f) Result : Complies / Does not comply
Done By: Checked By:
c. Calibration of Auto Injector by Linearity Measurement. (By Using FID with Capillary column):
Solvent Name | Batch No. | Calibration Standard No. | Purity in % | Use Before |
Tetrahydrofuran | ||||
Dimethylformamide |
(a) Standard stock solution: Weigh accurately _______(1.0 g) of Tetrahydrofuran (THF) standard in ______(50ml) volumetric flask containing 10 ml of DMF, dilute up to mark with ________DMF.
Standard / Sample solution Preparation: Take _______(2.0 ml) of stock solution in ______(50ml) volumetric flask containing 10 ml of DMF, dilute up to mark with ________DMF.
(b) Chromatographic condition :
Parameter | Test Conditions | Applied Conditions |
Column No. | ||
Column | AT-5 (30m X 0.53mm X 5.0µ) or Equivalent | |
Oven temperature | 100°C – 2 min. – @ 30°C/min.-250°-1min. | |
Detector temperature | 270°C – Flame Ionization detector (FID) | |
Carrier gas | Nitrogen | |
Head pressure | 30 kpa | |
Injection volume | 0.5µl, 1.0µl, 2.0µl, 5.0µl | |
Split ratio | 1:40 | |
Injector temperature | 200°C | |
Run time | 8 min. |
(c) Calibration procedure:
- Inject the sample with injection volume 0.5µl, 1.0µl, 2.0µl, 5.0µl.
- Inject each volume three times and record the observation in table.
- Calculate the %RSD of area at each volume. It should not be more than 10.0 %
- Plot the graph of injection volume vs. mean area of each volume, and calculate the correlation coefficient (r). It should not be less than 0.990.
(d) Observation Table
Injection Volume (µl) | Area of Tetrahydrofuran | Mean Area | % RSD (NMT 10.0 %) |
0.5 µl | |||
— | |||
— | |||
1.0 µl | |||
— | |||
— | |||
2.0 µl | |||
— | |||
— | |||
5.0 µl | |||
— | |||
— |
Correlation Coefficient _________________ (Limit: NLT 0.990)
(e) Result: Complies / Does not comply
Done By: Checked By:
d. Calibration of Auto Injector by Linearity Measurement. (By Using FID with Pack column):
Solvent Name | Batch No. | Calibration Standard No. | Purity in % | Use Before |
Isopropyl alcohol |
Water | Ultra Purified (Milli-Q) |
(a) Sample / Standard stock solution:
Weigh accurately _______(5.0 g) of Isopropyl alcohol standard in ______(50ml) volumetric flask containing 10 ml of water, dilute up to mark with ________water.
Take _______(3.0 ml of stock solution) in ______(50 ml) volumetric flask, dilute up to mark with ________water.
(b) Chromatographic condition:
Parameter | Test Conditions | Applied Conditions |
Column No. | ||
Column | Porapak Q – 80 / 100#, 2M, S.S. Column. | |
Carrier gas | Nitrogen | |
Flow rate | 50 ml /min | |
Detector | FID | |
Column temperature | 180°C | |
Injector temperature | 250°C | |
Detector temperature | 250°C | |
Injection volume | 0.5µl, 1.0µl, 2.0µl, 5.0µl | |
Run time | 15.0 min. |
(c) Calibration procedure:
- Set up the instrument as per the conditions.
- Inject the sample 0.5µl, 1.0µl, 2.0µl, 5.0µl each three times. (Calculate the area of IPA. Record the observation in below table.
- Calculate the %RSD of area for each injection volume. It should not be more than 10.0 %.
- Plot a linearity curve of Injection volume Vs corresponding mean area, using least square method. Calculate the correlation coefficient ® and record the observations in given below table.
(d) Observation Table:
Injection Volume (µl) | Area of Isopropyl alcohol | Mean Area | % RSD (NMT 10.0 %) |
0.5 µl | |||
— | |||
— | |||
— | |||
1.0 µl | |||
— | |||
— | |||
— | |||
2.0 µl | |||
— | |||
— | |||
— | |||
5.0 µl | |||
— | |||
— | |||
— |
Correlation Coefficient _______________ (Limit: NLT 0.990)
(e) Result: Complies / Does not comply
Done By: Checked By:
e. Calibration of TCD by Linearity Measurement. (By Using Pack column):
Solvent Name | Batch No. | Calibration Standard No. | Purity in % | Use Before |
Isopropyl alcohol | ||||
Acetone |
(a) Standard preparation:
Prepare the different Linearity preparations by mixing of IPA and Acetone (internal standard) as specified in the below table.
Linearity Preparation | Volume of IPA (ml) | Volume of Acetone (ml) IS |
1 | ___________(1 ml) | ___________(2 ml) |
2 | ___________(2 ml) | ___________(2 ml) |
3 | ___________(3 ml) | ___________(2 ml) |
4 | ___________(4 ml) | ___________(2 ml) |
5 | ___________(5 ml) | ___________(2 ml) |
(b) Chromatographic condition :
Parameter | Test Conditions | Applied Conditions |
Column No. | ||
Column | 10% PEG, 2 meter x 1/8” diameter, mesh rang 80-100 | |
Carrier gas | Helium / Hydrogen | |
Flow rate | 45 ml /min | |
Column temperature | 55°C | |
Injector temperature | 100°C | |
Detector temperature | 160°C | |
Current | 80 mA | |
Injection Volume | 1 µl | |
Run time | 10 min. |
(c) Calibration procedure:
- Set up the instrument as per the conditions.
- Inject each Linearity preparation in triplicate; calculate the mean area ratio of IPA to Internal standard Acetone.
- Calculate the %RSD of triplicate area ratio of IPA to Internal standard Acetone. It should not be more than 10.0%.
- Plot the graph of Concentration in ml Vs Corresponding mean area ratio of IPA /Acetone (IS) and calculate the Correlation coefficient (r) and it should not be less than 0.990.
- Record the all observations in the following table.
(d) Observation Table:
Volume of IPA (ml) | Area of IPA X | Area of Acetone Y | Area Ratio X/Y | Mean Area Ratio and % RSD |
1.0 ml | ||||
— | — | |||
— | — | |||
2.0 ml | ||||
— | — | |||
— | — | |||
3.0 ml | ||||
— | — | |||
— | — | |||
4.0 ml | ||||
— | — | |||
— | — | |||
5.0 ml | ||||
— | — | |||
— | — |
Correlation Coefficient ____________________ (Limit: NLT 0.990)
(e) Result: Complies / Does not comply
Done By: Checked By:
Remarks: The instrument is calibrated & qualified / Out of calibration & not qualified for use.
— | Calibrated By | Checked By | Approved By |
Signature | |||
Date | |||
Name | |||
Department | Quality Control | Quality Control | Quality Control |