Calibration Protocol of Polarimeter

Instrument Calibration Protocol of Polarimeter

InstrumentPolarimeter
Instrument ID. No. 
Ref. SOP No. 
Model No.                        P-2000
Calibration frequencyQuarterly – 5 days
Date of calibration               
Next due date 

1. Angle of rotation of Sucrose solution :

Weigh and transfer accurately 10 g of previously dried Sucrose AR to a 100 ml volumetric flask, dissolve and dilute with water and mix (10 %).

Make the different concentrations of Sucrose solution 20%, 30%, 40% and 50% in the same manner like 10 %. 

Balance ID: __________________.

Sucrose AR, Make: _________________, Valid up to :________________.

10 % w/v Sucrose AR Solution: _______________ g Sucrose AR  –> 100 with water

20 % w/v Sucrose AR Solution: _______________ g Sucrose AR  –>100 with water

30 % w/v Sucrose AR Solution: _______________ g Sucrose AR  –>100 with water

40 % w/v Sucrose AR Solution: _______________ g Sucrose AR  –> 100 with water

50 % w/v Sucrose AR Solution: _______________ g Sucrose AR  –> 100 with water

Temperature: 25 ± 0.5 °C

Angle of Rotation

ConcentrationObs.1Obs.2Obs.3Obs.4Obs.5AverageLimits
10 %      6.643 – 6.687
20 %      13.261 – 13.349
30 %      19.864 – 19.996
40 %      26.442 – 26.618
50 %      33.004 – 33.226

Remarks: Calibration Satisfactory/ Not Satisfactory

 Prepared ByChecked ByApproved By
Name     
Signature     
Name   
Department   