Instrument Calibration Protocol of Ultrasonic Sonicator with chiller
| Instrument | Ultrasonic Sonicator with chiller |
| Instrument ID. No. | |
| Ref. SOP No. | |
| Model No. | USB 50L |
| Calibration frequency | Half yearly – 15 days |
| Date of calibration | |
| Next due date | |
Calibration Record of Ultrasonic Bath & Ultrasonic Bath with chiller
| Ultrasonic Performance Meter ID | Calibrated done on | Calibrated Due on |
| | | |
| Sr.No | Location | Frequency (KHZ) (33 KHz ±3 KHz) | Power (%) (55 % ±5 % ) | Remarks |
| | Left Side corner (1) | | | |
| | Left Side corner (2) | | | |
| | Center (5 ) | | | |
| | Right Side corner (3) | | | |
| | Right Side corner (4) | | | |
Remark: Calibration Satisfactory/ Not Satisfactory
| | Prepared By | Checked By | Approved By |
| Name | | | |
| Signature | | | |
| Department | | | |